ACTION
|
RATIONALE
|
| 1. Check that
patient has a patent IV, and that the
defibrillator, emergency cart and appropriate
medications are available. |
1. This is an
emergent procedure. Venous access, emergent
equipment and drugs have a high probability of
use. |
| 2. Counsel
patient and family and allow for ventilation and
questions; obtain consent (time permitting). |
2. Time is needed
for coping; relieves anxiety; promotes
cooperation. |
| 3. Wash and dry
insertion site; clip hair if necessary. |
3. Prepare site
for dressing. |
| 4. Obtain vital
signs and ECG rhythm strip prior to insertion.
Connect to 12 lead EKG and continuously monitor
before, during and after. |
4. Establish
baseline parameters. This will assist in
monitoring for changes and complications. |
| 5. Prepare
according to central line insertion checklist. |
5. IAW Infection
Control Manual. |
| 6. Prep selected
site IAW central line insertion checklist. |
6. Reduce risk of
infection. |
| 7. Anesthetize
the area locally. |
7.. Reduce pain
from insertion. |
8. Prepare the
external temporary generator:
- Insert
new battery.
- Turn
the mA to 6.
- Turn
rate control to ten beats above the
patients rate.
- Turn
sensitivity dial fully clockwise.
|
8. The rate set
above the patients a rate will suppress the
patients natural pacemaker. |
| 9. Connect the
alligator clamp, one end to the V1 EKG lead and
the other to the distal or negative electrode
wire of the pacemaker catheter. |
9. This electrode
serves as an exploratory intracavitary lead and
resulting EKG waveform helps to ascertain
electrode position. |
| 10. Monitor
patient while physician inserts pacing electrode. |
10. The EKG
pattern is distinctive when the pacing electrode
is in the vena cava, right atrium and right
ventricle. |
CONFIRMING
POSITION:
The following
EKG findings indicate electrode position:
Location:
- Vena
Cava
- Right
Atrium
- Coronary
Sinus
- Right
Ventricular
- (Out
flow tract)
- Right
Ventricular Apex
|
Lead V1:
- Small
inverted p
- Tall
biphasic p
- Positive
QRS with rabbit ears
- Negative
QRS
- LBBB,
Negative QRS
|
Lead II, III,
AVF
- Same
- Same
- Positive
or neg QRS
- Positive
QRS
- Left
Axis/ pattern of injury (elevated ST)
|
Portable
Chest X-ray is required to confirm placement.
NOTE:
Insertion of a pacemaker during a cardiac arrest (asystole)
requires connecting the pacing wire to the generator and
setting the stimulation threshold to maximum or 20 mA,
and an arbitrary rate of 70 bpm. The electrode is then
advanced blindly while one scans the EKG for evidence of
ventricular capture.
POST
INSERTION PROCEDURE:
ACTION
|
RATIONALE
|
| 1. After contact
between pacing electrode and the right
ventricular endocardium is established, assist
the physician in connecting to pacemaker
generator. Set according to physicians
orders. Obtain a 12 Lead
EKG.
For
all Methods of Temporary Pacing
A.
Determine the stimulation threshold:
- Set
pacing rate above patients
intrinsic rate.
- Gradually
decrease output (mA) from 20 mA until
capture is lost.
- Gradually
increase output (mA) until capture is
established. This is the stimulation
threshold.
- Set
output (mA) at least 1.5 to 2 times
higher than the stimulation threshold.
This output setting is sometimes referred
to as the maintenance threshold.
B.
Determine the sensitivity threshold:
- Set
rate at least 10 beats per minute below
patients intrinsic heart rate.
- Set
sensitivity control to most sensitive
setting (fully demand or lowest numerical
setting). Sensing indicator light should
flash with each intrinsic R wave.
C. Set
the pacemaker rate, output (mA), and sensitivity
(demand or asynchronous), as prescribed or as
determined by threshold testing.
D.
Assess rhythm for appropriate pacemaker function:
- Capture:
is there a QRS complex for every
ventricular pacing artifact?
- Rate:
is the rate at or above the pacemaker
rate if in the demand mode?
- Sensing:
does the sensitivity light indicate that
every QRS complex is sensed?
|
1. The physician
sets the parameters; must generate a written
order for its use. Establish a
baseline.
The
output dial regulates the amount of electric
current (mA) that is delivered to the myocardium
to initiate depolarization. Ventricular pacing (stimulation
threshold) should be established at less than 1
mA output whenever possible. The maintenance
threshold is set at 1.5 to 2 times above the
stimulation threshold to allow for increases in
stimulation threshold without loss of ventricular
capture.
Sensitivity
threshold is the level at which intrinsic
ventricular activity is recognized by sensing
electrodes. For demand pacing, the sensitivity
must be measure and set.
Sensitivity
threshold is set at maximum and is lowered only
if pacer is sensing inappropriately. If set
too high, it will result in sensing P or T as
an R wave; if set too low, it results in
asynchronous pacing (nonsensing).
Determined
by patients response. A-V interval (similar
to intrinsic PR interval) should be set for
optimal ventricular filling, usually between 150
to 250 ms.
ECG
tracing should reflect appropriate response to
pacemaker settings if pacemaker is functioning
properly.
|
| 2. Ensure the
pacing wire is secured as it exists the
introducer; check all connections as well as
battery and control settings at least q.4 hrs and
document. |
2. Prevent
migration of wires; these connects are easily
disconnected and may result in cardiac arrest.
Protects setting from being inadvertently being
altered. |
| 3. Maintain site
care per Infection Control Manual SOP on central
lines. |
3. This is a
central line with potential for sepsis. |
| 4. Keep the pulse
generator dry and the controls protected from
mishandling. |
4. Moisture will
cause pacemaker malfunction; improperly set
controls can cause cardiac irritability resulting
in dysrhythmias and cardiac arrest. |
5. Protect the
patient from electromicroshock and
electromagnetic interference (EMI):
- Cover
exposed wires with gloves or tape.
- Enclose
pulse generator in rubber glove.
- Wear
rubber gloves when handling exposed wires.
- Avoid
any nurse-patient contact with electrical
apparatus.
- Check
for ungrounded electrical equipment.
- Keep
dressing over wires dry and intact when
not in use.
|
5.
Electromicroshock and EMI can cause the pulse
generator to fail. |
| 6. Monitor
patient for complications. Restrict patient
mobility depending on insertion site. |
6. All invasive
procedures have a percentage of complications
that can be prevented by an astute nurse. |
| 7. Document
location, type of pacing, pacing mode, stimulus
threshold, sensitivity setting, pacing rate and
intervals, intrinsic rhythm, and percent of
pacing. Post sample tracings. |
7. Proper
documentation assists in the care of the patient
and early recognition and prevention of
complications. |
Alll comments and questions
about content at this site should be sent to
Nurse Bob
There have been
Visitors to this website.
Return to Nurse
Bob's® Page
This page last updated 07/25/10
|